Remove the shipping foam insert (P/N CAP-122) from the plate area.

Often caused by a worn pivot point, a bent spindle, or improper sample volume. Refer to the manual's maintenance section to verify spindle runout.

The CAP 2000+ can be connected to a PC using the optional . This software allows you to: cap 2000 viscometer manual link

Essential steps to perform a calibration check using viscosity standards to ensure accuracy.

Select the rotational speed (RPM) and start the measurement. The unit will display centipoise (cP) or mPa·s. III. Maintenance Remove the shipping foam insert (P/N CAP-122) from

The main screen shows viscosity in your chosen unit (Poise, cP, Pa·s, or mPa·s), percent torque, temperature, spindle number, and actual speed. If the torque is below 10 % of full scale, the display flashes (under‑range condition). If the torque exceeds 100 %, “EEEE” is displayed (over‑range).

To find the manual link for the CAP 2000 Viscometer, you can try the following: The CAP 2000+ can be connected to a PC using the optional

: Always lift the unit by the base console or column ; never lift it by the handle or head.

Before calling technical support, the manual’s "Troubleshooting" chapter can help you resolve: